Publication date: Available online 16 February 2019
Source: Talanta
Author(s): Hugo Cunha-Silva, M. Julia Arcos-Martinez
Abstract
The World Health Organization considers iodide deficiency diseases (IDD) to be a public health problem. The main indicator to access IDD is urinary iodide, since approximately 90% of the ingested iodide uses this clearance path, with urine being a preferable target for the analysis.
In this work, two screen-printed carbon electrode (SPCE) based sensors were developed to determine iodide by using only a single drop of sample. A first approach based on a SPCE proves to selectively determine iodide through the control of the cathodic stripping voltammetric (CSV) parameters. However, this strategy exhibits a gap in determining trace iodide concentrations, which is improved by modifying the working electrode surface with a chitosan coating. The performance of this new CS/SPCE-based sensor was compared with that of the previous SPCE-based sensor, showing improved iodide determination sensitivity. A limit of detection of 1.0 ×10−8 M and a linear analysis range of 0.15 to 500 µM were achieved with this sensor.
The application of both sensors to real-life samples found values close to those determined by the standard Sandell-Kolthoff spectrophotometric method, proving them to be powerful analytical tools for iodide determination in different kinds of samples, including biological matrices.
Graphical abstract
from A via a.sfakia on Inoreader http://bit.ly/2S8RZYF