Abstract
Purpose
We aimed to determine optimal window settings for poly-energetic (PolyE) and virtual mono-energetic images (MonoE) derived from abdominal angiographic studies on a novel dual-layer spectral detector CT (SDCT) system.
Methods
From 50 patients, SDCT datasets PolyE and MonoE at 70 and 40 keV levels were reconstructed and best individual window width and level (BI-W/L) manually assessed. Through regression analysis, the so-called optimized individual (OI-W/L) values were obtained. Subjective image quality parameters and vessel diameters were measured to determine influences of different W/L settings.
Results
Image noise was lower and attenuation and contrast-to-noise ratio were higher in MonoE compared to PolyE (all p ≤ 0.002). Mean BI-W/L values for PolyE, 70, and 40 keV were 637/284, 647/291, and 1568/691, respectively. Mean OI-W/L values were 631/276, 628/286, and 1516/667, respectively. Compared to standard settings, all adjusted W/L settings varied significantly and yielded higher subjective scoring. No between-group differences were found between manually adjusted and mathematically calculated W/L settings.
Conclusion
PolyE and MonoE from abdominal angiographic SDCT studies require appropriate W/L settings especially at low energy reconstruction levels. Individual adjustment reaches the best image quality but is time consuming. From our data, predefined W/L settings of 640/280 (PolyE/MonoE 70 keV) and 1570/690 (MonoE 40 keV) as a non-individualized starting point for abdominal angiographic studies from the novel SDCT system are suggested.
from # All Medicine by Alexandros G. Sfakianakis via Alexandros G.Sfakianakis on Inoreader http://ift.tt/2tIVInm
via IFTTT
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου
Medicine by Alexandros G. Sfakianakis,Anapafseos 5 Agios Nikolaos 72100 Crete Greece,00302841026182,00306932607174,alsfakia@gmail.com,