Αρχειοθήκη ιστολογίου

Τετάρτη 16 Αυγούστου 2017

Biomechanical Measurement Error Can Be Caused by Fujifilm Thickness: A Theoretical, Experimental, and Computational Analysis

This is the first study to quantify the measurement error due to the physical thickness of Fujifilm for several material combinations relevant to orthopaedics. Theoretical and experimental analyses were conducted for cylinder-on-flat indentation over a series of forces (750 and 3000 N), cylinder diameters (0 to 80 mm), and material combinations (metal-on-metal, MOM; metal-on-polymer, MOP; metal-on-bone, MOB). For the scenario without Fujifilm, classic Hertzian theory predicted the true line-type contact width as , where is compressive force, is cylinder diameter, is cylinder length, and are cylinder and flat Poisson's ratios, and and are cylinder and flat elastic moduli. For the scenario with Fujifilm, experimental measurements resulted in contact widths of for MOM tests, for MOP tests, and for MOB tests, where is compressive force and is cylinder diameter. Fujifilm thickness error ratio showed a nonlinear decrease versus cylinder diameter, whilst error graphs shifted down as force increased. Computational finite element analysis for several test cases agreed with theoretical and experimental data, respectively, to within 3.3% and 1.4%. Despite its wide use, Fujifilm's measurement errors must be kept in mind when employed in orthopaedic biomechanics research.

from # All Medicine by Alexandros G. Sfakianakis via Alexandros G.Sfakianakis on Inoreader http://ift.tt/2w0LzUo
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Medicine by Alexandros G. Sfakianakis,Anapafseos 5 Agios Nikolaos 72100 Crete Greece,00302841026182,00306932607174,alsfakia@gmail.com,

Αναζήτηση αυτού του ιστολογίου

! # Ola via Alexandros G.Sfakianakis on Inoreader