Representing the dosimetric impact of deformable image registration errors

Deformable image registration (DIR) is emerging as a tool in radiation therapy for calculating the cumulative dose distribution across multiple fractions of treatment. Unfortunately, due to the variable nature of DIR algorithms and dependence of performance on image quality, registration errors can result in dose accumulation errors. In this study, landmarked images were used to characterize the DIR error throughout an image space and determine its impact on dosimetric analysis. Ten thoracic 4DCT images with 300 landmarks per image study matching the end-inspiration and end-expiration phases were obtained from 'dir-labs'. DIR was performed using commercial software MIM Maestro. The range of dose uncertainty (RDU) was calculated at each landmark pair as the maximum and minimum of the doses within a sphere around the landmark in the end-expiration phase. The radius of the sphere was defined by a measure of DIR error which included either the actual DIR error, mean DIR error...