The purpose of this work was to design and manufacture a 3D range-modulator for scanned particle therapy. The modulator is intended to create a highly conformal dose distribution with only one fixed energy, simultaneously reducing considerably the treatment time. As a proof of concept, a 3D range-modulator was developed for a spherical target volume with a diameter of 5 cm, placed at a depth of 25 cm in a water phantom. It consists of a large number of thin pins with a well-defined shape and different lengths to modulate the necessary shift of the Bragg peak. The 3D range-modulator was manufactured with a rapid prototyping technique. The FLUKA Monte Carlo package was used to simulate the modulating effect of the 3D range-modulator and the resulting dose distribution. For that purpose, a special user routine was implemented to handle its complex geometrical contour. Additionally, FLUKA was extended with the capability of intensity modulated scanning. To validate the...